Basic Operating Principle
Deadweight Testers (DWT) are the primary standard for pressure measurement. There are three main components of this device: a fluid (oil) that transmits the pressure, a weight and piston used to apply the pressure, and a connection port for the gauge to be calibrated.
The dead weight tester also contains an oil reservoir and an adjusting piston or screw pump. The reservoir accumulates oil displaced by the vertical piston during calibration tests when a large range of accurately calibrated weights are used for a given gauge. The adjusting piston is used to make sure that the vertical piston is freely floating on the oil. Please see How a Dead Weight Tester Works for a detailed description of the working principle of the device.
Calibration Basics
To carry out tests or calibrate a pressure gauge with the dead weigh tester(DWT), accurately calibrated weight masses (Force) are loaded on the piston (Area), which rises freely within its cylinder. These weights balance the upward force created by the pressure within the system:PRESSURE = FORCE/AREA = W/A |
Note:
if weights are in pounds (lbs) units and the area in inch square, then the calculated pressure unit is in Pounds per square inch(PSI).
If the weights are in kilogram (kg) units and the area of piston in meters square, then the calculated pressure [P = (W*G)/A, G = gravity in m/s2] unit is in N/m2 or pascal.
During calibration, the system is primed with liquid from the reservoir, and the system pressure is increased by means of the adjusting piston. As liquids are considered incompressible, the displaced liquid causes the piston to rise within the cylinder to balance the downward force of the weights.
Calibrating a Pressure Gauge with the Dead Weight Tester:
To calibrate a pressure gauge with a dead weight tester, set up the device on a level, stable workbench or similar surface as shown in the diagram below:
Proceed with the calibration according to the following steps:
Step 1:
Connect the pressure gauge to the test port on the dead weight tester as shown in the diagram above. Ensure that the test gauge is reading zero, if not correct the zero error and ensure that the gauge is reading zero before proceeding with the calibration exercise.
Step 2:
Select a weight and place it on the vertical piston
Step 3:Step 1:
Connect the pressure gauge to the test port on the dead weight tester as shown in the diagram above. Ensure that the test gauge is reading zero, if not correct the zero error and ensure that the gauge is reading zero before proceeding with the calibration exercise.
Step 2:
Select a weight and place it on the vertical piston
Turn the handle of the adjusting piston or screw pump to ensure that the weight and piston are supported freely by oil.
Step 4:
Spin the vertical piston and ensure that it is floating freely.
Step 5:
Allow a few moments for the system to stabilize before taking any readings. After system has stabilized, record the gauge reading and the weight.
Step 6:
Repeat steps 2 through 5 for increasing weights until the full range or maximum pressure is applied to the gauge and then decreasing weights until the gauge reads zero pressure. Calculate the error at each gauge reading and ensure that it is within the acceptable accuracy limits.
If you are doing a five point calibration, then increasing weights should be added corresponding to 0%, 25%, 50%, 75%, and 100% of the full range pressure of the pressure gauge. And for decreasing pressure you proceed in the order 100%, 75%, 50%, 25%, 0%.
For pressure gauges with less accuracy specifications, calibration at the points: 0%, 50% and 100% will suffice.
After calibration, your data can be recorded in a table in this manner:
Upscale Readings:
% Input | Weights,W | DWT Pressure (W/A)* | Test Gauge Pressure | Error |
0 | ||||
25 | ||||
50 | ||||
75 | ||||
100 |
*DWT Pressure = W/A, if W is in lbs, and A in square inch then DWT Pressure is in PSI(pounds per square inch). However, if W is in kg and A in square meters, then :
DWT Pressure = (W*G)/A, Where G = gravity in meters per seconds square(m/s2) and the DWT Pressure is in N/m2 or Pascal
Downscale Readings:
% Input | Weights,W | DWT Pressure (W/A)* | Test Gauge Pressure | Error |
100 | ||||
75 | ||||
50 | ||||
25 | ||||
0 |
At each pressure reading, the absolute error is calculated thus:
The absolute error at each point should be within the acceptable accuracy limits of the gauge.
If the gauge error is in % span proceed as follows to calculate the error:
The error in % span should be within the acceptable accuracy limits
otherwise the calibration will have to be repeated to correct the
errors.
If the pressure gauge error is in % FSD(Full Scale Deflection), proceed as follows to calculate the error:
Equations and factors are given on the certificate to adjust for any variations in these environmental conditions.
Always refer to the documentation for the Dead Weight Tester to ensure that for maximum accuracy, the necessary calibration correction factors are applied to any reading from the device.
reading. Due to the significant change in gravity throughout the world (about 0.5%), ensure that the tester in your possession has been manufactured with the specification of your local gravity, otherwise you may have to apply the correction for the calibrated gravity.
To correct for gravity use:
Absolute Error = DWT Pressure – Test Gauge Pressure |
If the gauge error is in % span proceed as follows to calculate the error:
Span = Maximum pressure – minimum pressure
%Error = [(DWT Pressure – Test Gauge Pressure)/Span]*100 for each pressure gauge reading.
|
If the pressure gauge error is in % FSD(Full Scale Deflection), proceed as follows to calculate the error:
% Error = [(DWT Pressure - Test Gauge Pressure)/FSD]*100 |
Correction Factors:
The deadweight tester has been calibrated to the Gravity, Temperature, and Air Density stated on the calibration certificate right from the laboratory.Equations and factors are given on the certificate to adjust for any variations in these environmental conditions.
Always refer to the documentation for the Dead Weight Tester to ensure that for maximum accuracy, the necessary calibration correction factors are applied to any reading from the device.
Gravity Correction
Gravity varies greatly with geographic location, and so will the deadweight testerreading. Due to the significant change in gravity throughout the world (about 0.5%), ensure that the tester in your possession has been manufactured with the specification of your local gravity, otherwise you may have to apply the correction for the calibrated gravity.
To correct for gravity use:
True Pressure = [(Gravity (CS))/(Gravity(LS))]*P(Indicated)
Where:
P(Indicated) = Pressure indicated by gauge being calibrated Gravity(CS) = Gravity at Calibration Site Gravity (LS) = Gravity at Laboratory Site |
Temperature Correction
Temperature and Air Density variations are less significant than gravity. Variations should be corrected for when maximum accuracy is required.To correct for Temperature variation use:
True Pressure = P(Indicated) [1+ {T(DWTCT) – T(OT)}*{ΔP/100}]
Where:
P(Indicated) = Pressure indicated by gauge being calibrated T(DWTCT) = Dead Weight Tester calibrated temperature in the laboratory T(OT) = Operating temperature at calibration site ΔP = Percentage pressure change per unit temperature change |